Aizah, Hana Silmi (2023) Pengaruh faktor eksposi terhadap dosis radiasi dan kualitas citra pesawat radiografi umum pada pemodelan phantom thoraks. Undergraduate thesis, Universitas Islam Negeri Maulana Malik Ibrahim.
|
Text (Fulltext)
19640065.pdf - Accepted Version Available under License Creative Commons Attribution Non-commercial No Derivatives. Download (2MB) | Preview |
Abstract
INDONESIA:
Salah satu pemanfaatan cahaya pada alat medis adalah dalam penggunaan radiasi sinar – x di bidang radiologi. Pesawat radiografi umum memiliki dua parameter yang mempengaruhi dosis dan kualitas citra yang akan dihasilkan, yaitu tegangan tabung (kV) dan arus waktu (mAs). Tegangan tabung dan arus waktu yang optimal mampu membuat gambar citra yang dihasilkan tampak jelas, akan tetapi semakin besar nilai dari tegangan tabung dan arus waktu yang digunakan maka akan semakin tinggi radiasi yang diterima oleh pasien. Diperlukan nilai tegangan tabung dan arus waktu yang sesuai sehingga dosis yang dihasilkan aman bagi pasien dan gamar citra yang dihasilkan memiliki kualitas yang baik. Penelitian menggunakan pesawat radiografi umum di dua rumah sakit umum yang ada di kota malang dengan memvariasikan nilai tegangan tabung dan arus waktu sehingga didapatkan nilai dosis keluaran alat yaitu DAP (Dose Area Product) dan nilai parameter kualitas citra yang meliputi CNR (Contrast to Noise Ratio), SNR (Signal to Noise Ratio), dan GM (Gray Matter) Conspicuity. Jenis phantom yang digunakan merupakan phantom akrilik. Pengeksposan phantom dilakukan dengan standar pemeriksaan thorax dengan memposisikan bagian permukaan phantom di bawah mesin sinar – X. Sinar – X yang menembus bagian phantom akan diproyeksikan menjadi gambar yang terdapat pada kaset radiografi dengan file berekstensi DICOM. Data citra yang telah didapatkan dapat dianalisis menggunakan beberapa software seperti ImageJ dan Matlab. Untuk mengubah ekstensi file dari DICOM ke JPG dan maka dapat menggunakan software ImageJ. Sedangkan software Matlab digunakan untuk menganalisis lebih lanjut mengenai analisis dosis yang didapatkan dan kualitas citra yang dihasilkan dengan menggunakan menu Graphical User Interface (GUI). Pada saat tegangan tabung divariasikan dan arus waktu konstan, nilai optimal pada rumah sakit 1 didapat pada faktor eksposi sebesar 56 kV 8 mAs dengan nilai DAP sebesar 35,66 mGy.cm2, CNR sebesar 2,093, SNR sebesar 24,323, dan GM Conspicuity sebesar 0,176. Sedangkan pada rumah sakit 2 didapat pada faktor eksposi sebesar 50 kV 8 mAs dengan nilai DAP sebesar 22,46 mGy.cm2, CNR sebesar 5,033, SNR sebesar 27,966, dan GM Conspicuity sebesar 0,236. Pada saat arus waktu divariasikan dan tegangan tabung konstan, nilai optimal pada rumah sakit 1 didapat pada faktor eksposi sebesar 20 mAs 48 kV dengan nilai DAP sebesar 37,08 mGy.cm2, CNR sebesar 5,127, SNR sebesar 27,380, dan GM Conspicuity sebesar 0,184. Sedangkan pada rumah sakit 2 didapat pada faktor eksposi sebesar 14 mAs 48 kV dengan nilai DAP sebesar 21,21 mGy.cm2, CNR sebesar 5,763, SNR sebesar 27,232, dan GM Conspicuity sebesar 0,293.
ENGLISH:
One of the uses of light in medical devices is in the use of x-ray radiation in the field of radiology. General radiography equipment has two parameters that affect the dose and image quality to be produced, namely tube voltage (kV) and time current (mAs). The optimal tube voltage and time current are able to make the resulting images appear clear, but the greater the value of the tube voltage and time current used, the higher the radiation received by the patient. It is necessary to have the appropriate tube voltage and current time values so that the resulting dose is safe for the patient and the resulting image has good quality. The study used public radiographs at two public hospitals in Malang city by varying the tube voltage and time current values so that the output dose value of the device was obtained, namely DAP (Dose Area Product) and image quality parameter values which included CNR (Contrast to Noise Ratio). , SNR (Signal to Noise Ratio), and GM (Gray Matter) Conspicuity. The type of phantom used is an acrylic phantom. Phantom exposure is carried out by standard thoracic examination by positioning the surface of the phantom under the X-ray machine. X-rays that penetrate the phantom section will be projected into an image contained on a radiographic cassette with a file extension DICOM. Image data that has been obtained can be analyzed using several software such as ImageJ and Matlab. To change the file extension from DICOM to JPG and then you can use the ImageJ software. While the Matlab software is used to further analyze the dose analysis obtained and the resulting image quality using the Graphical User Interface (GUI) menu. When the tube voltage is varied and the time current is constant, the optimal value at hospital 1 is obtained at an exposure factor of 56 kV 8 mAs with a DAP value of 35.66 mGy.cm2, CNR of 2.093, SNR of 24.323, and GM Conspicuity of 0.176. Whereas in hospital 2 the exposure factor was 50 kV 8 mAs with a DAP value of 22.46 mGy.cm2, CNR of 5.033, SNR of 27.966, and GM Conspicuity of 0.236. When the time current is varied and the tube voltage is constant, the optimal value at hospital 1 is obtained at an exposure factor of 20 mAs 48 kV with a DAP value of 37.08 mGy.cm2, CNR of 5.127, SNR of 27.380, and GM Conspicuity of 0.184. Whereas in hospital 2 the exposure factor was 14 mAs 48 kV with a DAP value of 21.21 mGy.cm2, CNR of 5.763, SNR of 27.232, and GM Conspicuity of 0.293.
ARABIC:
يمكن ملاحظة استخدام الضوء في القطاع الصحي من خلال استخدام العديد من أدوات الفحص الطبي كوسيلة لدعم التشخيص. من استخدامات الضوء في الأجهزة الطبية استخدام الأشعة السينية في مجال الأشعة. تحتوي معدات التصوير الشعاعي العامة على معلمتين تؤثران على الجرعة وجودة الصورة المراد إنتاجها ، وهما جهد الأنبوب (kV) والوقت الحالي (mAs). الجهد الأمثل للأنبوب وتيار الوقت قادران على جعل الصور الناتجة تبدو واضحة ، ولكن كلما زادت قيمة جهد الأنبوب وتيار الوقت المستخدم ، زادت كمية الإشعاع التي يتلقاها المريض. من الضروري أن يكون لديك جهد الأنبوب المناسب وقيم الوقت الحالي بحيث تكون الجرعة الناتجة آمنة للمريض وتكون الصورة الناتجة ذات جودة جيدة. استخدمت الدراسة الصور الشعاعية العامة في مستشفيين عامين في مدينة مالانج من خلال تغيير قيم جهد الأنبوب والوقت الحالي بحيث تم الحصول على قيمة جرعة الإخراج للجهاز ، وهي DAP (منتج منطقة الجرعة) وقيم معلمات جودة الصورة التي بما في ذلك CNR (نسبة التباين إلى الضوضاء) ، و SNR (نسبة الإشارة إلى الضوضاء) ، و مؤشر GM (المادة الرمادية). نوع الشبح المستخدم هو شبح أكريليك. يتم إجراء التعريض الضوئي الوهمي عن طريق الفحص الصدري القياسي عن طريق وضع سطح الشبح تحت آلة الأشعة السينية ، وسيتم عرض الأشعة السينية التي تخترق القسم الوهمي في صورة موجودة على شريط تصوير إشعاعي بامتداد DICOM. يمكن تحليل بيانات الصورة التي تم الحصول عليها باستخدام العديد من البرامج مثل ImageJ و Matlab. لتغيير امتداد الملف من DICOM إلى JPG ثم يمكنك استخدام برنامج ImageJ. بينما يتم استخدام برنامج Matlab لإجراء مزيد من التحليل لتحليل الجرعة التي تم الحصول عليها وجودة الصورة الناتجة باستخدام قائمة واجهة المستخدم الرسومية (GUI). عندما يتنوع جهد الأنبوب ويكون الوقت الحالي ثابتًا ، يتم الحصول على القيمة المثلى في المستشفى 1 عند عامل تعريض 56 كيلو فولت 8 مللي أمبير بقيمة DAP تبلغ 35.66 مللي غراي.سم2، كان CNR 2.093 ، و SNR 24.323 ، و GM Conspicuity كان 0.176. بينما في المستشفى 2 وجد أن عامل التعرض كان 50 كيلو فولت 8 مللي أمبير بقيمة DAP تبلغ 22.46 ملي جرام / سم.2و CNR 5.033 و SNR 27.966 و GM Conspicuity 0.236. عندما يتغير الوقت الحالي ويكون جهد الأنبوب ثابتًا ، يتم الحصول على القيمة المثلى في المستشفى 1 عند عامل تعريض 20 مللي أمبير 48 كيلو فولت بقيمة DAP تبلغ 37.08 مللي غراي.سم2، كان CNR 5.127 ، وكان SNR 27.380 ، وكان مؤشر GM هو 0.184. بينما في المستشفى 2 وجد أن عامل التعرض كان 14 مللي أمبير 48 كيلو فولت مع قيمة DAP 21.21 ملي غراي.سم2و CNR 5.763 و SNR 27.232 و GM Conspicuity 0.293.
Item Type: | Thesis (Undergraduate) |
---|---|
Supervisor: | Mulyono, Agus and Syarifah, Umaiyatus |
Keywords: | thoraks; dosis; citra; radiografi umum; thorax; dose; image; general radiography; الصدر .جرعة؛ صورة؛ التصوير الشعاعي العام |
Subjects: | 02 PHYSICAL SCIENCES > 0202 Atomic, Molecular, Nuclear, Particle and Plasma Physics > 020201 Atomic and Molecular Physics 02 PHYSICAL SCIENCES > 0299 Other Physical Sciences > 029901 Biological Physics 02 PHYSICAL SCIENCES > 0299 Other Physical Sciences > 029903 Medical Physics |
Departement: | Fakultas Sains dan Teknologi > Jurusan Fisika |
Depositing User: | Hana Silmi Aizah |
Date Deposited: | 23 Jun 2023 13:25 |
Last Modified: | 23 Jun 2023 13:37 |
URI: | http://etheses.uin-malang.ac.id/id/eprint/51682 |
Downloads
Downloads per month over past year
Actions (login required)
![]() |
View Item |